Editverse Free Calculator
Measurement Uncertainty Calculator (ISO 15189 GUM)
Estimate the measurement uncertainty of a laboratory result — combined and expanded uncertainty following GUM and ISO 15189 principles — with the exact R command behind the result. A full guide follows below. Validated against R 4.3.3.
In Simple Words
Measurement uncertainty (MU) quantifies the doubt around a laboratory result — the range within which the true value is expected to lie. ISO 15189 requires accredited laboratories to estimate it, and the GUM (Guide to the Expression of Uncertainty in Measurement) framework provides the standard method.
A practical estimate combines the main sources of variation — chiefly long-term imprecision from internal quality control and any bias uncertainty — into a combined standard uncertainty, then multiplies by a coverage factor (usually 2) to give the expanded uncertainty, the interval that captures about 95% of the spread around the result.
When to Use Measurement Uncertainty
Estimate MU for quantitative measurands as part of ISO 15189 accreditation, and to help clinicians judge whether a change between two results is real or within analytical noise. Reporting MU alongside results supports sounder interpretation.
Use a defensible data source — long-term internal quality control and external quality assessment — rather than a single short experiment, following the practical ISO/TS 20914 approach. Relate the expanded uncertainty to your sigma metric and quality goals. For agreement between methods rather than uncertainty of one method, see Bland-Altman. Unsure which framework applies? Our Test Selector can help.
Expected: one row per record, name, value, type
Drop a CSV here or click to choose. Parsed in your browser: nothing is uploaded.
Expected: one row per record, name, value, type
u <- c(0.12, 0.20, 0.05/sqrt(3), 0.10/sqrt(6)) # standard uncertainties
uc <- sqrt(sum(u^2))
k <- {k}
c(combined = uc, expanded = k*uc)User guide
When to use. Build a GUM uncertainty budget: convert each component to a standard uncertainty and combine them by root-sum-of-squares.
Assumptions. Components are independent (no covariance terms) and the measurand is a sum of the contributions with sensitivity coefficients of 1.
Common mistakes. Do not mix expanded and standard uncertainties. Include correlated or non-unit-sensitivity terms by hand before entering them.
Worked example
Click Load example data above, then Run Analysis to see a fully worked result with figure, formula and interpretation for this test.
References
- JCGM 100:2008. Evaluation of measurement data — Guide to the expression of uncertainty in measurement (GUM).
- Eurachem/CITAC Guide CG 4. Quantifying Uncertainty in Analytical Measurement. 3rd ed. 2012.
Worked Examples
Example 1 — From QC to expanded uncertainty
A measurand has a long-term internal QC imprecision of 2.5%. Treating that as the dominant component gives a combined standard uncertainty of about 2.5%, and an expanded uncertainty (coverage factor 2) of about 5%. A result of 100 units carries an uncertainty of roughly ±5 units.
Example 2 — Adding a bias component
If a known bias correction itself carries uncertainty, that component is combined in quadrature with the imprecision. For example, 2.5% imprecision combined with 1.5% bias uncertainty gives a combined uncertainty of about 2.9%, and an expanded uncertainty near 5.8%.
Assumptions to Check
A quantitative, stable measurand. GUM-style MU estimation applies to numeric results with definable sources of variation, not to qualitative tests.
Representative long-term data. The estimate should draw on long-term internal quality control and external quality assessment, capturing routine, not best-case, performance.
Correctly combined components. Independent uncertainty components are combined in quadrature (root-sum-of-squares); adding them directly overstates the uncertainty.
Common Mistakes
Using a single short precision run
One small experiment underestimates real-world variation. Long-term QC data give a realistic combined uncertainty.
Adding components linearly
Independent components combine in quadrature, not by simple addition; adding them inflates the estimate.
Forgetting bias uncertainty
If a bias correction is applied, its own uncertainty belongs in the budget; omitting it understates the true uncertainty.
Confusing standard and expanded uncertainty
The expanded uncertainty is the combined standard uncertainty multiplied by the coverage factor; reporting one as the other misstates the interval.
Frequently Asked Questions
What is the difference between standard and expanded uncertainty?
Combined standard uncertainty is the root-sum-of-squares of the components; expanded uncertainty multiplies it by a coverage factor (usually 2) to give an approximately 95% interval.
Why does ISO 15189 require measurement uncertainty?
Reporting MU lets clinicians judge whether a difference between results is analytically meaningful and supports transparent, accredited reporting.
What data should I use?
Long-term internal quality control and external quality assessment data, following the practical ISO/TS 20914 approach, rather than a single short study.
Does the result match R?
Yes. The calculator combines components in quadrature and applies the coverage factor, and shows the R command so you can reproduce the combined and expanded uncertainty.
Related Calculators
References
- Padoan A, Antonelli G, Aita A, Sciacovelli L, Plebani M. An approach for estimating measurement uncertainty in medical laboratories using data from long-term quality control and external quality assessment schemes. Clin Chem Lab Med. 2017;55(11):1696–1701. PMID 28245184.
- Padoan A, Sciacovelli L, Aita A, Antonelli G, Plebani M. Measurement uncertainty in laboratory reports: a tool for improving the interpretation of test results. Clin Biochem. 2018;57:41–47. PMID 29548924.
- ISO/TS 20914:2019. Medical laboratories — Practical guidance for the estimation of measurement uncertainty. Geneva: ISO; 2019. (See also JCGM 100:2008, GUM.)
Need This Analysis Done for Your Manuscript?
Our team builds ISO 15189 measurement-uncertainty budgets — combined and expanded uncertainty from QC and EQA data — with reproducible R output.